Reports 1-1 of 1 Clear search Modify search
AdV-INJ (Baffles for INJ benches (SIB 1, SIB 2, EIB and IMC cavity))
derossi, tringali, gosselin, spinicelli, bulten - 17:53 Thursday 14 March 2024 (63611) Print this report
INJ noise hunting shift on scattered light - LB and EIB noise injections

 

Following the investigation in the entry #63491, this morning we perfomed a set of noise injections on both laser bench and external injection bench.
We started with the injection on laser bench by means of shaker installed few days ago #63577.

LB NOISE INJECTIONS
ITF locked

DAC channel: ENV_NOISE_CEB_EEroom

Freq. line 17.9 Hz, A=0.1 V, 09:59:55-10:13:57 UTCFreq. line 8 Hz, A=0.7 V, 10:37:20-10:42:46 UTCFreq. line 8 Hz, A=0.6 V, 10:50:58-10:57:20 UTCSecond part of the shift ITF unlocked (INJ standalone)Quiet = 15:35:00 (5’)  Freq. line 17.9 Hz, A=0.1 V, 15:40:20-15:48:00 UTCFreq. line 8 Hz, A=0.7 V, 15:49:51-15:58:36 UTCFreq. line 8 Hz, A=0.6 V, 16:00:04  - 16:06:01 UTCColored noise (1-15) Hz: 16:18:27  - 16:23:39 UTC

EIB NOISE INJECTIONS

ITF unlocked (INJ standalone)

(previous EIB injections #61713)

We injected a sine of 0.5Hz frequency (Henk Jan adjusted the cfg file to have the right amplitude by just enabling sine noise button)

gps start DOF amplitude duration
16:24:26 z 10 um 3 min
16:28:08 y - -
16:33:31 x - -
16:37:46 tx 10 urad -
16:41:26 ty - -
16:44:45 tz - -

The analysis of these injections will follow.