AdV-COM (AdV commissioning (1st part) )
Allocca, Chiummo, Degallaix - 12:08 Wednesday 05 July 2017 (38435) Print this report
Comment to SSFS noise injections for CMRF measurments and OptChar tasks (38427)

Here some times from last night. All times are local times


SSFS noise injection

22:10 (5 min), noise injection to SSFS. Coherence mainly about 1 kHz.
Change the spectral shape of the injected noise to inject larger noise toward low frequency

22:30 (5 min), better, noise injection visible from 100 Hz. Try to increase the amplitude of the noise
Unlock, design a new filter with still larger amplitude in the low frequency band

22:37:30 (5 min) noise injection, slighlty better

23:46:50 Fmod enable

New filter shape for the noise injection: band pass centered around 30-40 Hz, not highly selective so we will still have noise injected outside the transmission band

00:04 (5 min) noise injected, quite good, well visible in h from 20 Hz up to high frequency

00:12:40 (5 min), increase the amplitude, best injection of the night, this data should be used

00:20:30 (2 min), increase again the amplitude of the noise, coherence not better in high frequency


Arm cavities free swinging and lock / unlock

00:54:20 North arm free swinging, scanning speed may be too low

01:05:00 Series of lock / unclock and at the same time series of free swinging

Problem with the logic of boolean and Metatron not cooperative. Remote help from Diego.

02:00 Go back to low noise, launch of the script for magnetic noise injection

 