Reports 1-1 of 1 Clear search Modify search
Detector Operation (Operations Report)
berni - 22:52 Thursday 16 November 2023 (62476) Print this report
Operator Report - Afternoon shift

The activity of the shift was again ITF recovery; activity concluded.

ITF left in LOCKED_ARMS_IR.

 

SUSP

NI missing data at 20:18 UTC.

 

Software

Bacnet server restarted at 20:15 UTC because it was providing flat channels.

Images attached to this report
Comments to this report:
masserot - 7:49 Friday 17 November 2023 (62477) Print this report

Since the 2023-11-15-13h04m47-UTC, there is no more missing data from Sa_WE . But the SC_NI missing data remains present.

Below the report of SUSP_Fb  server related to these events

  • 2023-11-15-12h41m10-UTC>INFO...-[TfbSourceManager::Clean] source Sc_NI (70:v07) has stopped -> waiting for data...
  • 2023-11-15-13h04m47-UTC>INFO...-[TfbSourceManager::Clean] source Sa_WE (812:v01) has stopped -> waiting for data...
  • 2023-11-15-13h25m42-UTC>INFO...-[TfbSourceManager::Clean] source Sc_NI (70:v07) has stopped -> waiting for data...
  • 2023-11-15-15h40m02-UTC>INFO...-[TfbSourceManager::Clean] source Sc_NI (70:v07) has stopped -> waiting for data...
  • 2023-11-15-17h04m52-UTC>INFO...-[TfbSourceManager::Clean] source Sc_NI (70:v07) has stopped -> waiting for data...
  • 2023-11-15-19h21m11-UTC>INFO...-[TfbSourceManager::Clean] source Sc_NI (70:v07) has stopped -> waiting for data...
  • 2023-11-16-08h17m54-UTC>INFO...-[TfbSourceManager::Clean] source Sc_NI (70:v07) has stopped -> waiting for data...
  • 2023-11-16-18h39m47-UTC>INFO...-[TfbSourceManager::Clean] source Sc_NI (70:v07) has stopped -> waiting for data...
  • 2023-11-16-19h05m29-UTC>INFO...-[TfbSourceManager::Clean] source Sc_NI (70:v07) has stopped -> waiting for data...
  • 2023-11-16-20h18m20-UTC>INFO...-[TfbSourceManager::Clean] source Sc_NI (70:v07) has stopped -> waiting for data...

 

Boschi, Passuello - 17:00 Tuesday 21 November 2023 (62523) Print this report

This morning a newer boot image was uploaded on Sc_NI board that is connected to the DAQ (PSDf in slot #1, serial #43148). The updated image contains a memory register change that should fix the data transmission issue.