Here attahed are acoustic noise transfer functions and projections computed with the injections performed on March 26 and on March 28:
- Figure 1 and first attached file: noise projection and TF in CEB hall with SQZ engaged
- Figure 2 and second attached file: noise projection and TFin CEB hall with SQZ shutter closed
- Figure 3 and 3rd attached file: noise projection and TF in Detection LAB with SQZ engaged
- Figure 4 and 4th attached file: noise projection and TF in Detection LAB with SQZ shutter closed
The difference with standard LN3+SQZ condition and that with SQZ shutter closed is a bit reduced projection and TF around 50Hz and 160Hz
Also attached is the matlab file to read the TF file and plot the data.