Reports 1-1 of 1 Clear search Modify search
AdV-COM (AdV commissioning (1st part) )
Montanari, Fiori, Soldani - 14:52 Thursday 20 July 2017 (38709) Print this report
Air conditioning switch off

Below a summary with all gps about the switch off of the air conditioning plus photon calibrator:

- Between 09:53LT and 10:04LT switches off in WEB
    boiler                                                Time: Jul20,17-09:53:05 -> GPS: 1184572401
    both chillers                                     Time: Jul20,17-09:53:52 -> GPS: 1184572448        
    both water pumps                           Time: Jul20,17-09:55:14 -> GPS: 1184572530
    UTA air conditioning                        Time: Jul20,17-09:56:34 -> GPS: 1184572610
    SWEB fans                                       Time: Jul20,17-09:59:16 -> GPS: 1184572772
    photon calibrator and driver            Time: Jul20,17-10:04:58 -> GPS: 1184573114
- 10:03LT wifi switched off (in all experimental areas)
- Between 10:15LT and 10:22LT switches off in NEB
    boiler                                               Time: Jul20,17-10:15:38 -> GPS: 1184573754
    both chillers                                    Time: Jul20,17-10:16:44 -> GPS: 1184573820
    both water pumps                           Time: Jul20,17-10:17:56 -> GPS: 1184573892
    UTA air conditioning                       Time: Jul20,17-10:19:23 -> GPS: 1184573979
    SNEB fans                                       Time: Jul20,17-10:22:15 -> GPS: 1184574151
- Between 10:29LT and 10:42LT switches off in CEB
    boiler                                              Time: Jul20,17-10:29:31 -> GPS: 1184574587
    hot water pump                              Time: Jul20,17-10:30:30 -> GPS: 1184574646
    cold water pump                            Time: Jul20,17-10:31:53 -> GPS: 1184574729
    chiller sx                                         Time: Jul20,17-10:33:14 -> GPS: 1184574810
    chiller dx                                         Time: Jul20,17-10:33:47 -> GPS: 1184574843
    DAQ air conditioning                      Time: Jul20,17-10:37:57 -> GPS: 1184575093
    Clean-Room air conditioning         Time: Jul20,17-10:41:05 -> GPS: 1184575281
    UTA hall air conditioning                Time: Jul20,17-10:42:18 -> GPS: 1184575354
- Between 10:51LT and 10:54LT switches off in MCB
    UTA air conditioning                      Time: Jul20,17-10:51:23 -> GPS: 1184575899                    
    chiller                                             Time: Jul20,17-10:52:29 -> GPS: 1184575965
    cold water pumps                          Time: Jul20,17-10:53:28 -> GPS: 1184576024
    heater                                            Time: Jul20,17-10:54:32 -> GPS: 1184576088
- Between 10:58LT and 11:01LT switches off in laser lab and ele-room
    fan coil (laser lab)                         Time: Jul20,17-10:58:54 -> GPS: 1184576350
    first splitter (inj ele-room)              Time: Jul20,17-10:59:39 -> GPS: 1184576395
    second splitter    (inj ele-room)     Time: Jul20,17-10:59:50 -> GPS: 1184576406
    UTA air conditioning (laser lab)    Time: Jul20,17-11:01:11 -> GPS: 1184576487
- Fan coil EERoom                            Time: Jul20,17-11:03:05 -> GPS: 1184576601    
- Fan coil DET electronic room         Time: Jul20,17-11:09:40 -> GPS: 1184576996
- UTA DET air conditioning               Time: Jul20,17-11:10:45 -> GPS: 1184577061

After that, we collected 5 minutes of data with ITF locked in LOW_NOISE_2 and then we switched all on without a timing

Comments to this report:
flood - 17:26 Thursday 20 July 2017 (38719) Print this report

Looking through the spectrogram of LSC_DARM we see a few lines disappear during the A/C switch off. Most of these 'lines' are farily broad small bumps in DARM. Correlating the time of the disappearance of lines with the switch OFF times. We can make the following correlations

 

  • @10:43LT 6.5Hz disappears (figure 1) -> 6.5Hz is a result of the UTA Hall A/C
  • @10:38LT 24.6Hz disappears (figure 2) -> 24.6Hz is a result of the DAQ A/C
  • @10:05LT 36.3Hz disappears (figure 3) -> 36.3Hz is the Photon Calibration Line 
  • @11:01LT 207Hz disappears (figure 4) -> 207Hz is a result of the second splitter (inj ele-room) //could also be the UTA A/C (laser lab), seems to be very high in frequency though

 

Asside from these lines there seems to be no other obvious lines that disappear. It is worth noting that in these are all relatively small bumps, with the exception of the Photon Calibration LIne. 

Images attached to this comment
mantovani - 12:05 Friday 21 July 2017 (38741) Print this report

The 207Hz disappeared at the switch off the INJ lab Air conditioning and it is probably associated to the M6 mount on the EIB see logentry 38640