Injection system (Injection bench installation)
pillant, genin, mantovani - 10:57 Monday 17 July 2017 (38640) Print this report
Optical mount resonances on the EIB

Friday we performed some tapping test on almost all the mirrors mount located on the optical path from the output of the Laser bench and SIB1.

The resonances and associated GPS of the tapping tests are listed below :

Mount GPS start GPS Stop Freq / DOF
M1 1184049487 1184049500 786Hz (hor)
M2 1184049531 1184049541 538Hz (ver) 529Hz (hor)
M3 1184049551 1184049563  weak
M4 1184049582 1184049596 350Hz(ver)
M5 1184049631 1184049644 410Hz
T2 1184049754 1184049764 547Hz
T3 1184049774 1184049781 491Hz
T4 1184049810 1184049818 441Hz 1019Hz
M6 1184049895 1184049903 201Hz  265Hz
M7 1184049931 1184049939 821Hz
M8 Before 1184049968 1184049975 226Hz  257Hz
M8 after 1184055900 1184055916 403Hz


The spectrum of the above listed injections are shown on the attached plots (from Figure 1 to 10, blue curve = exited purple curve = quite).

The mount of M8 has been changed, from Picture 11, to hardner the resonance. The comparison of the two noise injection on M8 is visible in Figure 12 (purple = old mount, blue = new mount).

The effect on the sensitivity is visible in figure 13 as a small improvement at 220Hz. It is worth to be noted that no worsening of the sensitivity is visible at the new resonant frequency @403Hz.