Reports 1-1 of 1 Clear search Modify search
AdV-INJ (Baffles for INJ benches (SIB 1, SIB 2, EIB and IMC cavity))
mantovani - 18:25 Thursday 28 July 2016 (34399) Print this report
B7 power dependence on input beam alignment

Today after the cavity has been aligned, by hitting the center of the NE and NI mirrors (using the dithering lines), and the B7 diodes have been centered the input beam has been aligned by changing the offset on the error signal of the PR vertical control.

The power drops of a factor 4 for a PR vertical displacement of ~ 100 um. The control of the beam in the cavity alignment can be done with the vertical position of PR.

Comments to this report:
mantovani - 18:26 Thursday 28 July 2016 (34401) Print this report

The missing plot

Images attached to this comment