Reports 1-1 of 1 Clear search Modify search
Injection system (General activities)
lagabbe, gosselin, zaza - 18:05 Wednesday 26 August 2026 (69654) Print this report
Chiller of the neovan electronic has been exchanged

The chiller of the neovan electronics stopped working at around 14:00 utc. Probably due to normal device aging.

This afternoon we exchanged the chiller with the spare that was in the optics lab. The serial number of the old chiller is on the pictures attached to this post.

PS: In the chiller water tank, we found some 'sand' which is probably some metal oxide and particle coming from the electronic devices and tubing.

Images attached to this report
Comments to this report:
lagabbe, zaza, spinicelli - 16:25 Thursday 27 August 2026 (69662) Print this report

This afternoon around 14h20 local, we increased the chiller temperature setpoint from 16 C to 17 C, and increased the water flow.

Images attached to this comment
spinicelli, zaza, lagabbe - 12:05 Friday 28 August 2026 (69665) Print this report

In addition to the two action reported, the "energy safe" option has been selected yesterday too (see fig. 1). 

Other than the usual 1Hz comb noise,  present with those chiller when the safe mode is not selected, we could see a sawtooth oscillation on T of the chiller and the neoVAN diodes, with a period of about 11min (see fig. 2).  
The oscillation went away as soon as we changed the operation mode.

Images attached to this comment