All the Acl servers (AclISC, AclLC, AclSBE, AclFCoplev, Acl) are running with the v1r23p17 version since the 2024-03-26-15h20mn-UTC .
The main new features are
- for the Cm commands use to set the ACL object parameters, the numeric parameters can be sent as integer or double .
- the ACL_SINECOMB_CH keyword to define a comb of lines
- the possibly to to delay the readout of the AclAdcCh objets.
The first trial was done the 2024-03-25 starting at 14H-LT
- all the Acl servers were restarting the Acl v1r22p17 : operation complete around 17h-LT
- upgrade of the INJ Acl server configurations
- Some tuning were done on the NEB and WEB Als servers to avoid missing ADCs sample by adjusting the timing sequence: the LOOP_DELAY was increased to 6us for NEB and 7us for WEB
- after the recovery of Injection system and the ITF alignment, we faced some troubles with the lock sequence
- some Cm messages sent by the Automation servers with the "load" or "apply' commands related to ACL_MATRIX_CH or ACL_SUM_CH objects were all rejected by the Acl servers
- To relock the ITF, the LSC_Acl, ASC_Acl, NEB_ALS_BPC, WEB_ALS_BPC and the SDB1_LC servers were restarted with the Acl v1r20p17 version . The ITF was relocked at LN3 . Unfortunatly , the SDB1_OMC server was not updated and the OMC lock sequence was not correctly achieved
- To fix the issue the Acl v1r23p17 was created and deployed on all the Acl servers