Here we give the times of the today switch off test, in sequence:
11:45LT _ stop of turbo's cooling fans of: PR ; BS ; PR-BS (these are the near ones to PR tower: on PR turbp, on BS turbo, on PR-BS link turbo, respectively)
12:45LT _ stop of the HVAC cooling fan of the 'remote pump' room
13:30LT _ stop of all the remaining active cooling fans of CEB turbos: IB; DT;SR;NI;WI (both types, industrial ones and 'home' type ones)
13:30LT _ restart of the HVAC cooling fan of the 'remote pump' room
16:30LT _ restart of all the turbo fans previously switched OFF.
Remark: since the PR; BS; PR-BS fans are likely affecting the ITF signals (forthcoming analysis), we have applied there a first seismic insulation support during
the restart operation itself (rubber feet + heavy supporting slab). To be improved during the next maintenance break.