Reports 1-1 of 1 Clear search Modify search
AdV-TCS (CO2 laser projector)
bersanetti, boldrini, bossilkov, corubolo, lumaca, mantovani, nardecchia, pinto, spinicelli, ruggi - 21:18 Thursday 17 September 2026 (69770) Print this report
TCS tuning part.2

This morning, we continued the tuning activity started yesterday (69761). 
The initial TCS settings are summarized in the table below.

 

STARTING POINT

After STEP 2 (logbook entry 69761)

CH [W]

INNER DAS [W]

OUTER DAS [W]

W

on the ITF

0.026 

0

0.90

on the pickoff 

0.160 

0

0.150

N

on the ITF

0.026

0.15 

0.60

on the pickoff 

0.160

0.024

0.098

The CARM NULL locks were limited by a vertical oscillation of the PR at around 11 Hz, which was causing the ITF to unlock. Paolo and Manuel took care of the issue and managed to solve it.

At some point, we also experienced a few unlocks due to the green losing lock. This issue was investigated and solved by Piernicola.

At this point, since the B1p fringe was already quite dark (~0.016 mW), we decided to try a common step on both DASes in an attempt to increase the sideband power. We therefore decided to increase both the WI and NI DAS powers by 20% at 14:42 UTC (STEP 3)

Unfortunately, shortly afterwards, the ITF unlocked due to an earthquake, which prevented further lock acquisition attempts for some time (earthquake at 14:19 UTC).

Once the earthquake transient had approximately settled, we noticed that the sideband powers in CITF had decreased by about 10%.

ISC then attempted two lock acquisitions, but both failed at the same step (CARM offset, step 3/3).

Therefore, before retuning the lock acquisition, we decided to also increase both the WI and NI CH powers by 20% (STEP 4, at 16:25 UTC).

The sideband increased but not recovering the level before the STEP 3. Some attempts were made to reach CARM NULL, but we did not manage to reach it. The ITF always unlocked at the same step (CARM offset, between step 2 and 3).

The final CO₂ laser configuration is summarized in the table below.

 

FINAL POINT 

After STEP 4

CH [W]

INNER DAS [W]

OUTER DAS [W]

W

on the ITF

0.031 

0

1.08

on the pickoff 

0.192 

0

0.18

N

on the ITF

0.031 

0.18

0.72

on the pickoff 

0.192 

0.029

0.118

 

 