gouaty, lunghini - 10:53 Tuesday 01 September 2026 (69677) Print this report
OMC demodulation phase and B1_PD3 offset adjusted

This morning, the OMC was not locking due to a mistuning of the OMC error signal demodulation phase, which was found to be wrong by ~90 deg.

The demodulation phase of B1_PD3 was adjusted to 3.4 rad (previous value was 1.85 rad). After reloading the external configuration at 07h45m54 utc, the OMC was able to lock on the TEM00 at first attempt.

Assuming that the demodulation phase for the B1_PD2 photodiode would be affected in the same way, we also adjusted the B1 demodulation phase from -0.6 rad to 0.97 rad. The new demodulation phases have been recorded in the main configuration file of process SDB1_OMC.

We also noticed that the offset of the B1_PD3 photodiode was quite large. The offset was retuned and we reloaded the configuration of SDB2_dbox_bench at 08h05m03-UTC (see attached figure).

Images attached to this report
Comments to this report:
mwas - 12:22 Tuesday 01 September 2026 (69678) Print this report

Figure 1. The adjustment today has essentially reversed the adjustment I had done on Sunday evening. When the interferometer was DOWN for a few hours on Sunday afternoon the offset has increased over a few hours. Once I did the offset adjustment, and relocked the arm the offset has decayed back on a time scale of 12 hours. Which made reversing the Sunday adjustment necessary. 

Figure 2. The offset on B1 PD3 is closly following the temperature changes on the bench, which is what one could expect. And the adjustment on Sunday happened to be at the peak of the temperature excursion following the issues with SDB2 on Sunday afternoon. https://logbook.virgo-gw.eu/virgo/?r=69669

Images attached to this comment