The up-left coil looking from the rear side has an issue with the shielding of its cableĀ https://logbook.virgo-gw.eu/virgo/?r=69386
Is that the same coil-magnet pair which is not responding to DAC actuation? Electrical checks should be done as a priority in any case, but if the up-left coil-magnet pair is the issue, then it makes it more likely that this is an electrical problem.
Plots showing the result of WI MIR actuation cheks are attached. If a voltage is generated by UL DAC (seen by the monitor inside the board), no effect is visible on the mirror orientation (fig 1). If the same is done with one of the other DACs, the rotation is well visible (fig 2).