We deployed the DAC1955 v3r3 firmware on almost all the DAC1955 boards used for the SDB1, SDB2 et EDB systems.
- rtpc13 : the 2 DAC1955 boards are updated
- Tolm output data flow reduced from 4.2MB/s to 0.42MB/s
- rtpc1 : 9 DAC1955 boards over the 13 are updated
- it remains the 2 related to the SQZ and the 2 driven by the EDB server
- Tolm output data flow reduced from 54.5MB/s to 20.6MB/s
Operations performed the 2026-06-31 between 2026-06-30-06h56m23-UTC and 2026-06-30-15h16m55-UTC
- all the loops were opened during the time of the operations
- and successfully closed at the end
Below the detailled list of the DAC1955 boards updated and the related ACL's server
- SDB2_LC and SDB2_SBE were already done for the 3 DAC1955 used
- SDB2_DAC_LC_LVDT_out : DAC1955_SN019 - DBOX_SN028 - SDB_EDB_dbox_rack : rtpc1
- SDB2_DAC_LC_LVDT_in : DAC1955_SN069 - DBOX_SN106 - SDB2_dbox_bench : rtpc13
- SDB2_DAC_SBE0 : DAC1955_SN022 - DBOX_SN027 - SDB_EDB_dbox_rack : rtpc1
- loops successfully closed
- EDB_PC (Phase Camera) : 1 DAC1955
- SDB2_DAC_EDB0 : DAC1955_SN023 - DBOX_SN027 - SDB_EDB_dbox_rack : rtpc1
- plot showing the DAC1955 input channels and the Phase camera related outputs
- SDB1_FI : 1 DAC1955
- SDB1_DAC : DAC1955_SN042 - DBOX_SN016 - SDB_EDB_dbox_rack : rtpc1
- we checks that the Tolm packect is correctly received by the DAC board
- SDB2_Quadrants : 1 DAC1955
- SDB2_DAC_galvo : DAC1955_SN101 - DBOX_SN105 - SDB2_dbox_bench : rtpc13
- we checks that the Tolm packect is correctly received by the DAC board: It remains to check that the galvo loops are correctly running
- SDB1_OMC, SDB1_Quadrants and SDB1_FAST_SHUTTER : 1 DAC1955
- SDB1_DAC_OMC_B5 : DAC1955_SN021 - DBOX_SN030 - SDB_EDB_dbox_rack : rtpc1
- plot showing the successfull OMC temperature stabilization and the presence of the PZT modulation line
- we checks the B5 quadrant VBias are OK
- It remains to check the FAST_SHUTTER is correctly running
- SDB1_LC : 3 DAC1955
- SDB1_DAC_LC_LVDT_out : DAC1955_SN012 - DBOX_SN028 - SDB_EDB_dbox_rack : rtpc1
- SDB1_DAC_LC_LVDT_in : DAC1955_SN025 - DBOX_SN016 - SDB_EDB_dbox_rack : rtpc1
- SDB1_DAC_MAR : DAC1955_SN048 - DBOX_SN030 - SDB_EDB_dbox_rack : rtpc1
- the loops were successfully closed but the TX and TY set points have a little bit shifted, the experts must check
- the EDB_DAC2 board installed bu not yet used has been updated too
- EDB_DAC2 : DAC1955_SN055 - DBOX_SN113 - SDB_EDB_dbox_rack : rtpc1
During the operations, we updated by msitake the firmware of one mezzanine(id 3) of a DBOX (SN23) used for the SPRL local control : we restored the correct firmware and after the SPRB LC were successfully closed too