After the successfull deploiement for the WEB NCAL , we continue to use this new firmware on WEB and NEB PCal system and for the NEB NCal one.
Below the list of the operations performed:
- WEB PCAL system
- Update the DAC1955-SN11 (WEB_PCAL_DAC) located in the DBOX-SN49 with the v3r3 release for the FPGA firmware and the DSP code
- Update the WEB_PCal_fast Acl server
- Update the WEB_PCAL_DAC part in the SWEB_dbox_rack DBox server and reconfigure it
- After these operations the WEB_PCAL loop was successfully close
- operations performed between 2026-05-05-06h49m50-UTC and 2026-05-05-07h27m54-UTC
- NEB NCAL and PCAL systems update
- Update the VPM Online configuration to use the DaqBox v17r9 for the SNEB_dbox_{bench,rack} servers
- Update the DAC1955_SN6 (NEB_DAC_NCAL0) ) located in the DBOX-SN118 with the v3r3 release for the FPGA firmware and the DSP code
- Update the DAC1955_SN18 (NEB_DAC_NCAL_TCS) located in the DBOX-SN119 with the v3r3 release for the FPGA firmware and the DSP code
- Update the DAC1955_SN121 (NEB_DAC_PCAL) located in the DBOX-SN35 with the v3r3 release for the FPGA firmware and the DSP code
- Update the NEB_NCal server configuration
- Update the NEB_PCal server configuration and the NEB_NNC_Tilt server configuration
- these 2 servers are sharing the same DAC1955_SN121 (NEB_DAC_PCAL) but each one sends a Tolm packet related to its used channels
- Update the SNEB_dbox_rack configuration to set the related DAC1955 boards configuration
- operations performed between 2026-05-05-08h28m20-UTC and 2026-05-05-09h39m12-UTC
- After these operations only the NEB NCAL loops were tested and all were successfully closed
- for the NEB_rtpc, the output data flow related to the DAC1955 boards was reduced from 29.4MB/s to 19.3MB/s