Reports 1-1 of 1 Clear search Modify search
Injection system (General activities)
spinicelli, vardaro - 15:41 Thursday 05 February 2026 (68634) Print this report
Crystal temperature change of the ML

After the change of the crystal temperature of the last summer, the main laser of the squeezer was too far from the Virgo ML (see here and there). 

In agreement with Marco, we lowered the Virgo ML temperature by ~0.45⁰ to try improving squeezing system stability, which seemed to be stable in July. PSL/INJ system relocked quickly after the operation (a check of the OLTF of the IMC didn't spot any problem).

The locking of the squeezer laser is ongoing.

We will check in the following days the stability of the PSL/INJ to be sure the new T setpoint is stable enough.

Comments to this report:
vardaro - 20:40 Thursday 05 February 2026 (68635) Print this report

Here the new set point of Squeezing PLL

PLL MAIN: 29668

PLL CC: 29640
PLL SC:  36026

 

Search Help
×

Warning

Error

The present report has been modified outside this window. Please check for its integrity in the main page.