Reports 1-1 of 1 Clear search Modify search
AdV-PAY (Payloads Control)
pinto, ruggi, bersanetti - 12:23 Wednesday 28 January 2026 (68592) Print this report
PR MAR actuaction balancing

Some misbalancing of the MAR actuation on PR has been noticed.

By looking at the data this occurred during the measurement performed on vertical DoF of PR F0, which triggered to move the payload with respect to the F7 in vertical of around -700 um to a bistable position of the anti-springs (see fig.1: Sc_PR_F7_LVDT_V).

This eventually caused a loss in the accuracy of the PRty alignment loop, see fig.2 in which a stretch of data have been taken in LN3 before the misbalancing (blue) and after (red).

By looking at the TF between MAR Zcorr and TYcorr we measured the needed coefficient to adjust the actuation balancing (fig.3). This adjustment has been done this morning. Once we will have a PR cavity locked we will check the outcome of the action.

Images attached to this report
Comments to this report:
pinto - 12:58 Wednesday 28 January 2026 (68594) Print this report

comparison in CARM null 1f of the PRty aligment loop accuracy before and after (blue and red) the balancing.

Images attached to this comment
Search Help
×

Warning

Error

The present report has been modified outside this window. Please check for its integrity in the main page.

Refreshing this page will move this report into drafts.

×