ITF found in relocking phase, LN3 automatically reachead at 7:05 UTC.
At 7:09 UTC I set Calibration mode to run the requested calibration:
- CALIBRATED_DF_DAILY
- inject_lsc.py
- inject_ssfs.py
- inject_asc_tes.py
At 8:23 UTC ITF in Commissioning mode to run the requested ENV noise injections; the ITF unlocked at 9:25 UTC while performing acoustic noise injections at WE.
The ITF relocked at the first attempt without any manual intervention; Science mode was set at 10:10 UTC but it unlocked again at 10:33 UTC.
After some failed attempts the ITF was again in Science mode at 11:56 UTC.
At 12:24 UTC ITC in Commissioning mode to complete the ENV noise injections (script had been changed by Maria to restart from WE acoustic injectio); the ITF unfurtunately unlocked at 12:29 UTC. At this point it was decided to skip the ENV injections.
The ITF relocked again without any manual intervention and Science mode was set at 13:24 UTC.
ITF left in Science mode with Autoscience ON.
ENV
During the shift there were periods of high wind activity.