Leaving SDB1 moved along Z axis by ~5 mm. In case of any problems
in the lock acquisition during the night the F0 X set point should be
moved back to its initial position, 2600 um X set.
This has improved the noise in LN2 by ~4Mpc, to be confirmed what it
does in LN3. In any case it doesnt' make it much worse as a few minutes after reaching LN3 the BNS range was 54Mpc.
Figure 1 shows the translation of SDB1 along the beam axis that makes the BNS range better, but the optical gain worse.
Figure 2 shows
Some more detailed notes on the measurements done today with the EDB OMC that were the initial motivation of the shift. And some measurements done on Tuesday at the beginning of the IPATSiA shift.
The EDB B1t photodiode DC and Audio calibration needs to be updated, it is still using the parameters from before the photodiode replacement last week.
----
Sep 30
Locked on 56MHz sideband, I think it is the upper sideband, but I am
not quite sure. And did adjusted of 100-200 step on EDB_OMC1_TX/TY
17:32 UTC (3 min) calm data after realignment
Started scan with same parameters as last time. frequency 0.002Hz,
amplitude 0.5 degree, offset 23.27 degree. This corresponds to a one
degree step in the starting point. I guess with the air flux in
reduced mode the working temperature is higher than in high air flux
as used last week during the tuning of the photodiode during work on
SDB2 with people present in the room for several days. So beginning of this scan will be highly non linear.
turned off EDB picomotors
------
Oct 2
The Bt1 photodiode Volt to Watt conversion factor have not been
updated after the PD swap. So the channels are no longer properly
calibrated. The DC channel is a factor 2 smaller than the Audio one. I
am not sure what is the absolute calibration either. In the analysis I
have multiplied the DC channel by 2 to match the Audio one, and
assumed the Audio one is correct, which gives a resonable level of
noise at high frequency that can be explained as shot noise.
Interferometer in LN2
15:43 UTC - lock on 56MHz USB, turn on pico-motors. Managed to almost
completely remove the beam jitter peaks at a few hundred Hz.
15:58 UTC (5min) 56MHz USB quiet data. The beam jitter peaks remained
low for the 5min of the measurement
turned off picomotor driver
when passing thurog 56MHz USB order 1 mode it was mostly vertical and
5 times lower in power than TEM00
16:10 UTC (5min) order 2 carrier - spoiled by saturation
noticed that B1t Audio saturates all the time on the order 2 and 3 modes
reduced DIFFp TX/TY lines to the LN3 amplitude (instead of LN2)
16:40 UTC (5min) order 3 carrier
reduced DIFFp TX line by another 25%
16:51 UTC (5min) order 2 carrier
17:02 UTC (5min) order 3 carrier
17:13 UTC (5min) order 4 carrier
17:24 UTC (5min) order 5 carrier
17:35 UTC (5min) order 6 carrier
17:53 UTC (5min) order 2 carrier
18:01 Moving SDB1 suspension, F0 X starting point 2600 um
this misaligns the EDB OMC, relocking on 56MHz USB TEM00
turning EDB picomotors on realigning EDB OMC in vertical as the
translation of SDB1 along the beam axis progress.
The optical gain is decreasing (OMC mode matching getting worse), but
the BNS range is clearly improving by several Mpc. The SDB1 B5
quadrant has the beam near the edge of the linear range so going
further is not reasonable. Stopping with F0 X at -2200, so 4.8mm total
motion according to F0 position calibration.
19:03 UTC (5min) order 2 carrier
19:21 UTC (5min) order 3 carrier
19:27 going to LN3 with EDB OMC remaining locked on order 3 carrier
mode. However the EDB OMC will get misaligned as the bench turns to
keep the SDB1 OMC aligned.