AdV-SGD (FDS commissioning)
bertolini, guo, spagnuolo, vandenbossche, vardaro - 18:36 Thursday 10 April 2025 (66550) Print this report
Recovery of filter cavity alignment with green

The filter cavity is again relocked with the green beam.

Preparatory work

  • This morning Luca checked the vacuum level around FCIM tank then open the gate valve
  • We recovered the vertical position of SQB1 and SQB2 benches
  • FCIM inverted pendulum was put in the same set points of July 2024 (last FC lock)

Realignment of the cavity

  • Before closing FCIM local control loop we checked the transmission of the green beam on the camera in transmission from the filter cavity FCEB_GR_Cam. We saw that the beam was too high and clipped thus we have to change the correction of EQB1_GR_M7 from 2 to -5 in order to see the beam completely on the camera and the beam centered also on FCIM external surface (toward SQB2)
  • We went to SQB2 and we checked on the viewport the presence of the back reflection
  • We left open FCIM _LC_MAR_T{X,Y} and we kicked the input mirror sending DC offset to the marionette. During this oscillation we saw the FCIM reflection passing randomly on SQB2
  • We looked the RDS data and we found a moment in wich the back reflection arrived to EQB1_GR__PD_DC . Thus we closed the LC with this set point and we saw immediatlely the reflection coming back up to EQB1
  • Once the retroreflection was coming back up to EQB1 we started to move the end mirror in order to find cavity flashes on FCEB_PD_DC photodiode in transmission
  • To do it we proceeded similarly as FCIM. This time with the LC closed we moved a lot TX and TY set point and a certain point we saw a small glitch in the transmission PD
  • With RDS data we went to the position of the glitch and we found the first flashes
  • As soon as we had flashes we aligned the cavity in the standard way but in order to maximize the transmission we had to move in vertical also EQB1_GR_M5 and M7 mirrors
  • Once the transmission was higher than 3V we engaged the lock and the autoalignment and it worked at the first attempt
  • The maximum value obtained with automation is 4V
  • We realized that EQB1_GR_M{5,7}_Y DOFs were saturiating. Thus we went int DET_LAB to discharge them

 

Comments to this report:
sorrentino - 14:48 Tuesday 22 April 2025 (66603) Print this report

Today around 12:20 UTC I restored the air covers on FCIM optical levers, see pictures

Images attached to this comment
vardaro - 15:43 Tuesday 22 April 2025 (66605) Print this report

After the covering of the FCIM optical levers I found SBE and LC loop out of range. I recoverd before the IP position loops

Than I had to close the Optical lever loop one by one

Finally the reflection of FCIM was not reaching anymore SQB1. With the help of ipcam61 I was able to find again the beam  back and relock the filtercavity