Plan for the day
- Test of angle of incidence and mirror loss on reflection bench of NE PCal
- Test of angle of incidence and mirror loss on reflection bench of WE PCal
- Test of NE_Rx at WE PCal
Test of the angle of incidence and mirror loss on NE reflection bench mirror
1- mirror loss measurement
Around 7h20UTC we opened the reflection bench and placed the WSV sphere in front of the M3 mirror to measure the beam right after the viewport.
At 7h25 UTC, with laser at 1.3 W we measured a power on WSV = 1.29882 +- 3.5e-4 W
At 7h36 we put WSV sphere at its initial position and measure a power on WSV =1.2966 +- 3.9e-4 W
Those measurement are giving a value of mirror loss of 0.17% (comparable to LAPP measurement of 0.12% and last measurement made on site in June 2023 of 0.12% also)
This seem to indicate that most of the discrepency between NE and WE are coming from the viewport rather than the M3 mirror.
2- Visual inspection of the viewport
In order to visually inspect the viewport on the reflection bench we opened the diaphragm and shined a flash light onto it, the inspection was not conclusive as it seem to be some minimal scratch and dust but not particularly in the middle of the viewport.
After this we tried closing the diaphragm as it was before.
A first measurement made at 7h52UTC gave us a value of power on WSV = 1.29576 +- 3.5e-4 W
We tried closing the diaphragm more at 7h55 UTC and measured a power on WSV slightly lower.
We put the diaphragm back at it initial setting at 7h58UTC.
At 8h05 UTC We closed NE Rx bench, with WSV at Rx spot and the laser at 1.3 W to redo a short calibration measurement for the photodiodes to check if touching the diaphragm changed the calibration.
Calibration AFTER playing with the Rx diaphragm. (PCal NE lines stopped)
THe background measurement is done using the same data than for the first calibration : 1427652018 for 3600s
The time period for this calibration measurement is from 10h10UTC to 13h39 : 1427710218 for 12540s
old gain | new gain | diff | old offset | new offset | |
Tx_PD1 | -0.775251 | -0.772392 | 0.37% | -0.000757 | -0.000768 |
Tx_PD2 | -0.858945 | -0.858746 | 0.023% | -0.000843 | -0.000861 |
There is no significant change from the first calibration performed on the 2025-04-03 before touching the diaphragm. The plot of the voltages ratio from WSV to the photodiodes is in plot1.
Test of angle of incidence and mirror loss on reflection bench of WE PCal
We arrived at WE building at 8h25, and stoped the laser and Rx power supplu at around the same time.
1- Test of Rx blue box value when not plugged but still powered
8h30 : 15 V power supply stopped (blue box not stopped)
~8h33 : Rx cable unplugged on Rx sphere
8h33 : blue box stopped, power supply switched on → WE_Rx_PD1_DC goes up to 3.6 W constant . ADC saturation?
8h36: blue box switched on, still constant value.
8h37 : blue box and power supply off . Value back to ~0
8h40 : Rx sphere plugged, then power supply on, and blue box on-> value ~0.
The same effect was seen later with WSV when retrieving it from NE, in plot 2 we can see three steps of power value, first when the laser was on at 1.3W reading around -3.5 V.
Then when the laser was turned off and the WSV blue box turned off with a value around 0V and finally when the cable was disconnected from the sphere where we have a high reading of around 3 V (see plot 3) similar to the Rx case seen before.
2- Measurement of mirror loss
During this test we noticed some glitches in the data, more investigation will be done but not mirror loss measurement were performed
Measurement of Rx_NE noise at WE
8h43 : laser back to 1.3 W.
8h45 : Stop PCal lines on WE.
10h45 : start of the noise level measurement of Rx_WE
10h47 : stoped laser and stop blue box of Rx_WE
10h50 : replace the Rx sphere of WE by the Rx sphere of NE to test the noise of Rx_NE.
10h58 -> 11h05 : start blue box, start laser and ctrl loop at 1.3 W. -> Channel PCAL_WE_Rx_PD1_DC is now reading the power using the Rx sphere from NE.
11h10 : WE_Rx sphere put back, laser switched on to 1.3 W
11h16 : put back calibration lines on NE and WE
In plot4 we can see the comparison of the noise level of the Rx_WE and Rx_NE sphere with the photodiodes as witness channels.
We see that the increased noise level for Rx_NE is also seen when reading it with Rx_WE electronics indicating that the higher noise is coming directly from the NE sphere itself.
Retrieving WSV from NE
Arrived at the NE building around 13h30 UTC, turned off laser around 13h35UTC
FInished removing WSV and its electronic at 14h05 UTC
Turned Rx and laser back on at 14h10 UTC