AdV-DET (Commissioning)
gouaty, mwas, bersanetti - 16:36 Tuesday 04 March 2025 (66300) Print this report
DET_MAIN stalled

This afternoon the DET_MAIN node got stalled in SHUTTER_CHECK_CLOSED.

We suspect that the cause of the problem was a change in the electronic offset on the B1_PD3 photodiode (correlated with the detectiion lab humidity control test performed during the maintenance).

We adjusted the B1_PD3 offset by +0.0002 mW at 15h22 utc. Then after, performing a successful SHUTTER_CHECK_OPEN (since the shutter was actually opened), we performed a SHUTTER_CHECK_CLOSED to restore the standard conditions of DET_MAIN.

Comments to this report:
mwas - 15:42 Wednesday 05 March 2025 (66301) Print this report

Figure 1 . Looking at the offset during the relock today the offset is well adjust with the power at zero when there is no power on the B1 PD3.