This morning, we prepared the setup for the acoustic injections at WEB, scheduled for Thursday. Below are the parameter values (in bold) for the acoustic injections using the VPM ENVnoise process:
********************************************************************************************************************
# nameOut - unit - rampTime - samplingFreq - density - seed
ACL_NOISE_CH noise_white "V" 5 SAMP_FREQ 1.
ACL_FILTER_CH noise_colored "V" 5 SAMP_FREQ noise_white 0.004 "flt3"
********************************************************************************************************************
These values have to be set to 0.0 and 0.04 for the automated CEB acoustic injection with the Metraton node ENV_MAIN:
********************************************************************************************************************
# nameOut - unit - rampTime - samplingFreq - density - seed
ACL_NOISE_CH noise_white "V" 5 SAMP_FREQ 0.0
ACL_FILTER_CH noise_colored "V" 5 SAMP_FREQ noise_white 0.04 "flt3"
********************************************************************************************************************
The next step will be implementing automated acoustic injections in the terminal buildings.