Yesterday morning, during the maintenance we checked the status of the neovan chiller. Despite a small increase in temperature in the last week, the chiller flow was quite low. We then decided to change the filter with a clean one and swap the water cables (thus inverting the flux). Thanks to this, we recover a much better flow (as a reminder, we do not have online data from the chiller), and a result we decreased the NEOVAN head temperature by more than 1⁰C (see fig. 1).
Since yesterday, the temperature is stable as well as the other ISYS channel (in particular, PSTAB rms improved quite a lot compared to the last days, see fig. 2 too).