gherardini - 6:57 Monday 11 November 2024 (65508) Print this report
Operator Report - Night shift
The ITF kept the lock all the shift, science mode stopped from 0:28UTC to 0:34UTC because of squeezing disengage, the squeezing was not able to relock most probably because of a change in the detection lab humidity (see plot), so I left the ITF without squeezing for the rest of the night.

- guard tours (UTC):
22:12 --> 22:40
0:06 --> 0:36
2:08 --> 2:36
4:11 --> 4:41
Images attached to this report
Comments to this report:
mwas - 15:14 Monday 11 November 2024 (65510) Print this report

Figure 1. After the squeezing was disengaged two bumps appeared in the specturm around 25Hz and 30Hz. These are related to the squeezer automatic alignment dither signals, and likely mean the residual B1p beam is misaligned on the squeezing benches when the squeezer is switched off.

Images attached to this comment
vardaro - 18:49 Wednesday 13 November 2024 (65525) Print this report

In order to reemove the lines I forced the switch off of the line in DOWN,
I re switch on the line in SQZ_INJECTING. I pressed load duirng the Troubleshooting