After the afternoon shift, I restored the TCS configuration settled on November, 17th 2023.
At 21:57 UTC PR CHRoCC VDAC 0.42 --> 0.08 (21600 sec ramp time) by the operator (see entry 62497).
At 22.28 UTC SR RH switched off via telnet. ( I try to use the VPM command, but it seems not to respond).
The settled CO2 actuator's powers are listed in the table below:
Sun Nov 19th 22.50 UTC | CH [W] | INNER DAS [W] | OUTER DAS [W] | |
W | on the ITF | 0.036 | 0 | 0 |
on the pickoff | 0.225 | 0 | 0 | |
N | on the ITF | 0.024 | 0.08 | 0.55 |
on the pickoff | 0.147 | 0.013 | 0.089 |
The checks have been performed between 22.32 UTC and 22.55 UTC.
In the end, all the CO2 actuators needed for CITF and lock acquisition flip mirrors have been opened.