Reports 1-1 of 1 Clear search Modify search
AdV-INJ (Suspended injection bench (SIB1) body)
gosselin, derossi - 22:48 Tuesday 19 September 2023 (61714) Print this report
Comment to SIB1 noise injections (61616)

We redid the noise injection in LN3

DOF GPS start UTC duration filter amplitude
quiet 21:45:00 5 min    
PR X 22:08:20     5e-9
" 22:11:08 3 min   5e-8
quiet 22:15:00 2 min    
PR Y 22:22:20 2 min   5e-8
" 22:22:35 2 min   1e-7
  22:24:05     1e-6 too high, we unlocked

For PR_X (ie SIB1 TY), we were sending more noise at high frequency and less at low frequency than last week (see the filter at the bottom of the entry), we saw many new structures in the range 50-200 Hz (see figure 1 and 2).
For PR_Y we were sending noise similar to what we did last week but it made the interferometer unlock.

# ASC Noise filter butterworth
ACL_FILTER_SET             ASC_noise_butter        1    1    10  12
ACL_FILTER_BUTTERWORTH     ASC_noise_butter    "lowpass"  2 3    200
ACL_FILTER_BUTTERWORTH     ASC_noise_butter    "highpass" 10 3    50

Images attached to this comment
Search Help
×

Warning

Error

The present report has been modified outside this window. Please check for its integrity in the main page.

Refreshing this page will move this report into drafts.

×