During the first part of the shift, we worked with the ITF in single bounce configuration to check the alignment of the B1 photodiodes, then to check the alignment and mode matching of OMC:
DET MAIN slow shutter automation has issues in single bounce. It fails because it checks the B1p photodiode status immediately after opening the B1p PD2 shutter, before it had time to be updated in the Metatron data that are a few seconds behind. Need to increase the delay of the check to see if it can fix the issue.
Retuned the OMC demodulation phase, was off by 2.5 radian.
Have seen correlation in OMC transmission with B5 quadrant, mainly vertical. Realliged OMC using picomotors, 200 steps in vertical first motor, 100 steps in horizontal first motor and about double that on the second motor (to get a beam shift).
Locked on 56MHz LSB, with 5 times slower scan (0.1 degree of amplitude with 0.1mHz). Observed 0.7mW in transmission of OMC. No light found on B1 PD1/PD2 at first attempt of shutter opening. On second trial B1 PD2 shutter opened. For B1 PD1 we had to try at least ~5 times before it opened. After that could open/close B1 PD1/PD2 shutters without problem.
08:12:30 UTC (2min) OMC locked on LSB 56MHz with no action. Maybe could analyze 56MHz RIN later with these data.
Checking B1 PD1/PD2 alignment by moving SDB1 in TY by ±500urad. Beam on B1 camera moves by ±0.8mm. At opposite edges of motion each photodiode losses 1% (one edge for PD2 and the other for PD1).
Moving in TX by ±100urad. Moving by ±0.2mm no clear impact on B1 PD1/PD2. More power on B1s PD1.
Realigned B1s PD1 using picomotor in reflection of OMC on SDB1.
08:34:30 - Up-going scan 0.1 degree with 1mHz frequency:
0.65uW - LSB 56MHz TEM00
0.105mW - carrier TEM00
0.65uW - USB 56MHz
0.95uW - carrier order 1 mode vertical
carrier order 2 mode saturated B1 PD1/PD2 and closed fast shutter
0.02mW on B1 PD1/PD2 for order 2 mode of USB 56MHz
08:43 down going scan with B1 PD1/PD2 closed:
0.05uW on order 2 mode of USB 56MHz
5.5uW on carrier order 2 mode
08:47 end of scan
08:53 UTC (3min) OMC locked on carrier light with no action. Correlation with SDB1 TX still visible, around ±1%, for ±50um on B5 QD2 V.
Adjusted SDB1 in Z by ~2mm (see Fig.1). Realligned a little bit the OMC with picomotors in H. Transmitted power clearly improved by ~5%.
09:44:40 (3min) scan after mode-matching and alignment (see Fig.2):
0.11mW on TEM00
0.001mW on order 1
0.0027mW on order 2
We improved the mode matching but the order 2 mode is still at 2.5% of TEM00. As we did not want to spend too much time on mode matching tuning (considering that the input beam will be better matched to the FP cavities next week) we decided to stop the mode matching tuning here and try to lock the OMC at dark fringe.
After Piernicola and Camilla restored a stable lock at dark fringe, we put a dark fringe offset LSC_DARM_SET=-0.2 and we scanned the OMC in these conditions (amplitude 0.12 deg, frequency 1 mHz), starting from 13h05m30 (see Fig.3). A fine analysis of this scan will be performed later, below is what we observed on the fly:
TEM00 of SB (56 LSB ?) at 6.5 mW.
Carrier TEM00 at 11 mW.
TEM00 of SB (56 USB?) at 15 mW.
Carrier first order mode at 20 mW.
Carrier second order mode at 20 mW.
Carrier third order mode at 8 mW.
Carrier fourth order mode at 16 mW.
We started a down-going scan but the ITF unlocked at 13h16m36 due to a large power on B1s (reaching the threshold of 43 mW that triggered the closing of SDB1 fast shutter).
After the ITF was relocked in dark fringe, we tried to the lock OMC but failed (Fig.4). A first attempt of OMC lock is shown in Fig.5 with the dark fringe offset put at -0.2. The power on the carrier TEM00 was fluctuating around 20 mW, which should have been OK. The scan parameters were amplitude=0.12 deg and frequency = 0.0003 Hz.
Another failed attempt of OMC lock with an increased dark fringe offset at -0.3 is shown in Fig.6.
Then we stopped trying to lock the OMC and tried to scan it around the carrier TEM00. Fig.7 shows the crossing of the carrier TEM00. What is strange is that the power seems to be reduced (no more than 7 mW in transmission) while the DARM offset is still at -0.3.
Further analysis has to be performed to better understand the causes of this two failed OMC lock attempts.
At the end of the shift, upon request from the commissioning coordinator, we brought bench the SDB1 bench at its initial longitudinal position in order to not risk any side effect on the Hartmann beam. Fig.8 shows that while the SDB1 bench is moving along Z, the B1p position on the camera is changing by ~0.8 mm. Note that the SDB1 drift control was not engaged during that time, which is probably the reason for this large alignment drift not compensated the alignment of SDB1.
Next time we will work on the OMC, we should start by checking the alignment and the mode matching again.