ITF found in Science Mode, switched to Calibration Mode at 5:20 UTC to start the Magnetic Injection.
Maintenance started at 6:00 UTC, here a list of the actions communicated to the Control Room:
6:05 UTC - 9:50 UTC - Grass cutting around Central Building Area ( N. Baldocchi );
6:10 UTC - 9:20 UTC - Standard Vacuum Refill ( VAC Team );
6:25 UTC - 10:05 UTC - Standard cleaning inside the experimental Area ( M. Ciardelli );
6:45 UTC - 7:25 UTC - Activity on suspension platform around NI and WI in preparation for Temperature sensor calibration for Etalon ( G. Sposito );
7:55 UTC - Weekly TCS Chillers and refill, only the WI unit required a small refill ( A. Magazzu );
8:06 UTC - 8:52 UTC - IMC Scan ( INJ Team );
9:00 UTC -11:05 UTC - WE UPS2 repairment ( M. D'Andrea );
9:25 UTC - 10:20 UTCÂ - Temperature sensor calibration for Etalon inside Clean Room ( V. Dattilo, G. Sposito );
Maintenance completed at 10:13 UTC, after some attempts the ITF was back in Science Mode from 11:01 UTC.
ITF left locked.
DAQ
SQZ_MAIN crashed at around 10:00 UTC after an unlock. Properly restarted.
EnvMoni
Weekly Magnetic Injection started at 5:20 UTC and completed at 6:02 UTC.
SBE
From 10:40 UTC to 10:50 UTC, due to a loss of power on B7_DC, I adjusted SNEB LC_TX. Value changed from -640 to -652.