TCS DAS works
Summary of the shifts 7-8-9/11/17
 
During these shift we worked on the CO2 benches for the fine tuning of the intensity distribution on the DAS beams (in the figure 1 the NI DAS in the CP position is shown).
 
Today we have turned on the WI RH to check the position of the TM center on the HWS maps:
RH ON-> 13.46 UTC
RH OFF->15.30 UTC
P_{dissipited)=2 W (1 W for each ring)
 
The position has not changed as you can see in figures 2 and 3.
 
During these shift we worked on the CO2 benches for the fine tubino of the intensità distribution on the DAS beams (see figure 1).
 
 
Today we have turned on the WI RH to check the position of the TM center on the HWS maps:
RH ON-> 13.46 UTC
RH OFF->15.30 UTC
P_{dissipited)=2 W (1 W for each ring)
 
THe position has not changed as you can see in figures 2 and 3.
Images attached to this report
Comments to this report:
schaaf, tacca - 17:20 Monday 20 November 2017 (40093) Print this report

Images obtained with phase camera while the WI RH is turned on.

1. figure attached shows the power on B4 and B1p as observed with the main diodes. The time at which the ring heater is turned on is indicated in red. The power leaking to the dark port clearly increases with the WI mirror heating up. The blue lines indicate the gpstimes at which the images shown in the following are taken. Note that the steady state is not reached for the image with ring heater on (and an unrelated unlock follows shortly after).

2. figure attached shows the power images on B4. The upper row is the situation with ring heater turned off, the lower once the ring heater is turned on for about 1 hour (as indicated in 1. figure). The images are averaged over 10 s. Slight changes in the sidebands are visible.

3. figure attached shows the power images on B1p. The images are again averaged over 10 s. Most noteworthy the 08 mode seems to reduce in intensity. A misalignment becomes visible.

4. figure shows the overlaps. The overlap of carrier with the 56 MHz sidebands is clearly reduced while the 6 MHz sidebands see much less of an effect. The overlap with the 6 MHz USB seems even to improve.

5. The sidebands and the carrier on B4 were fitted with an astigmatic Gaussian. The data is better described by the fit when the WI RH is turned on, i.e. the overlap of fit with data increases (see formula in figure). The astigmatism of the sidebands is reducing aswell (again formula in figure).