AdV-PAY (NE and WE Payloads)
ruggi - 20:35 Thursday 06 March 2025 (66317) Print this report
Measurement of electric charge on WE

A line at 19.9 Hz, amplitude 1V, has been injected on each WE MIR coil in configuration OPEN, which means connected to the electronics only on one side. In this way no current flow is generated by the injection, but only a voltage. During the injection, only one pair of WE coils was used to lock. The one not used for the lock and not used for the injection was totally disconnected, when the line was injected on UR and DL. Going to UL and DR, a large effect on the 50 Hz was induced in DISCONNECTED state, out of the range of the feed forward applied on WI, so the one not used was kept in the standard state (CLOSED in LN2).

The gps of the injections are:

gpsDL=1425315918;
gpsUR=1425316718;
gpsDR=1425318518;
gpsUL=1425319518;
gpsCLean=1425321068;

duration: 300s (even much more, in some cases)

The attached plot shows hrec in one case (the others are similar except for the impact on the 50 Hz). The amplitude of the injection was set in order to have a signal at the double frequency (39.8 Hz). That signal was small, but we decided to stop increasing the line because the effect at 19.9 Hz was already very large.

The same injections have been performed on NI. The amplitude of the line was 0.3 V. The three coils not involved in each injection were in standard state (LN);  the effect on the 50 Hz was large in any case.

The gps are:

NIUL=1425322018;
NIDL=1425322718;
NIUR=1425323818;
NIDR=1425324498;

300s are always available, but sometime the data are spoiled by a glitch, visible as a drop of the range. In order to select the better periods, the state of the injections are visible in the channels Sc_NI_MIR_VOUT_UL etc.
 


 

Images attached to this report
Comments to this report:
ruggi - 10:01 Friday 14 March 2025 (66372) Print this report

Yesterday the injection of a line at 19.9 Hz on each MIR coil in OPEN configuration has been performed on WI (0.3 V amplitude) and NE (1 V amplitude). gps are listed below. 300 s or more are available for each injection.

gpsCL=1425915218;
WIUL=1425916918;
WIDL=1425917918;
WIUR=1425919918;
WIDR=1425921368;
NEDL=1425922268;
NEUR=1425923618;
NEUL=1425924718;
NEDR=1425926118;
 

salvador - 11:55 Wednesday 02 April 2025 (66488) Print this report

Due to the use of a shaping filter in the coil driver boards of NE and WE, the injected voltage values differ from the settings. The injection values can be retrieve looking at the specific voltage channel of the coil.

Using the GPS of the injections, here are the values of the voltages as seen by the V1:Sc_XX_MIR_VOUT_YY (with XX: WE, WI, NE, NI and YY: DL, DR, UL, UR) channels:

Mirror Coil Vinj (V) 
NE DL 0.194
NE DR 0.195
NE UL 0.205
NE UR 0.202
NI DL 0.316
NI DR 0.316
NI UL 0.314
NI UR 0.317
WE DL 0.196
WE DR 0.195
WE UL 0.193
WE UR 0.192
WI DL 0.316
WI DR 0.314
WI UL 0.316
WI UR 0.316
salvador - 17:22 Thursday 03 April 2025 (66502) Print this report

Here are the extracted values of the voltage and charge of the mirrors extracted from the data for LSC_DARM and Hrec_hoft_20000Hz.

DARM:

Mirror Coil Qmir (pC) Vmir (V)
NE DL 87.8+-37.0 77.1+-32.5
NE DR 111.8+-71.6 98.2+-62.9
NE UL 101.4+-47.6 89.1+-41.8 
NE UR 84.0+-32.7 73.8+-28.7
       
NI DL 705.8+-125.8 619.9+-110.5
NI DR 486.8+-83.8 427.5+-73.6
NI UL 306.0+-269.4 268.8+-236.6
NI UR 445.2+-433.4 391.1+-380.7
       
WE DL 451.2+-176.4 396.3+-154.9
WE DR 764.8+-330.9 671.7+-290.7
WE UL 758.3+-306.9 666.0+-269.6
WE UR 791.5+-320.5 695.2+-281.5
       
WI DL 98.6+-20.5 86.6+-18.0
WI DR 31.2+-8.2 27.4+-7.2
WI UL 99.7+-17.4 87.6+-15.3
WI UR 106.7+-17.2 93.7+-15.1

 

Hrec:

Mirror Coil Qmir (pC) Vmir (V)
NE DL 259.5+-98.8 227.9+-86.8 
NE DR 303.9+-165.4 266.9+-145.3
NE UL 269.5+-129.4 236.7+-113.6
NE UR 215.8+-78.1 189.5+-68.6
       
NI DL 1861.8+-296.8 1635.3+-260.7
NI DR 1281.4+-263.2 1125.5+-231.1
NI UL 813.3+-722.9 714.3+-635.0
NI UR 1187.3+-1171.7 1042.8+-1029.1
       
WE DL 1182.4+-430.6 1038.5+-378.2
WE DR 1911.4+-663.2 1678.8+-582.5
WE UL 2076.4+-678.7 1823.8+-596.2
WE UR 2132.3+-959.9 1872.8+-843.1
       
WI DL 267.8+-57.9 235.2+-50.9
WI DR 83.0+-18.6 72.9+-16.4
WI UL 260.8+-43.6 229.1+-38.3 
WI UR 291.1+-50.7 255.6+-44.5

 

These were extracted using the ASD values for f=19.9 Hz and 2f=39.8 Hz. The difference between DARM and Hrec is explained by the difference in the calibration factor at the 2 different frequencies.

The error values were extracted using the sqrt of the variance of the ASDs used for averaging the data.