In this afternoon shift science mode stopped from 16:00UTC to 19:02UTC for daily calibration and commissioning (PRCL offset scan and ISC noise injections) and for troubleshooting from 14:45UTC to 14:48UTC to solve the problem with Hrec; the ITF unlocked during the commissioning and it relocked at the first attempt.
- guard tours(UTC):
17:40 --> 18:10
19:40 --> 20:10
ISC
- 18:24UTC: LSC injections (inject_lsc.py);
- 18:40UTC: ASC injections (inject_asc_test.py);
because of a lack of time injections on SSFS and PSTAB not done, it will be run tomorrow afternoon;
Other
from 14:07UTC to 14:45UTC we had a strange behavior of Hrec with the range going up and down randomly and the sensitivity too, it was because of a test done by Alain on a test version of Hrec that by mistake was writing in a wrong shared memory.