Reports 1-1 of 1 Clear search Modify search
AdV-TCS (CO2 laser projector)
Aiello, Fafone, Nardecchia, Rocchi, Ruggi - 23:29 Tuesday 21 November 2017 (40115) Print this report
DAS tests with P ITF @26.4 W

During this shift we tested some DAS configurations to check the effect on the ITF.

More details on the different steps will follow tomorrow. The  last test has been performed with P_in=125 mW and P_out=250 mW. This configuration seems close to the optimum working point.

We monitored the ITF behaviour with these DAS settings for about 3 hours. The ITF is stable, so we leave it locked with the DAS ON.

 

 

 

Comments to this report:
aiello, fafone, nardecchia, rocchi, ruggi - 17:37 Wednesday 22 November 2017 (40117) Print this report

Here we report the details of the DAS actuation tests during the yesterday shift. The same power has been applied on both benches.

Test Start (UTC) End (UTC) Power IN (mW) Power OUT (mW)
1 16:04 16:06 200 400
2 16:24 16:40 100 200
3 16:45 17:33 150 300
4 17:33 17:52 175 350
-- 18:04* 19:25* 100* 200*
5 19:40 -- 125 250

*during this measurement the WI laser changed its working point, so the data are not reliable (grey area in the attached plot).

In the white areas the DAS was OFF.

The selected working point (test 5, probably close to the best one) allowed to keep the ITF locked for the all night.