AdV-PSL (Installation on site)
derossi, pillant genin - 15:52 Monday 28 May 2018 (41594) Print this report
main laser chiller replacement

This morning, at around 08h55 UTC, we replaced the main laser chiller (see 1st plot).

The new chiller is from termotek, model P30210E-16750-2 (serial # 93612) and it has a higher flow speed. However, we checked the signals from the accelerometers and seismometers on the laser bench and we did't notice a worsening from the previous situation (see plot 2).

Images attached to this report
Comments to this report:
derossi, pillant, ciardelli, genin - 15:15 Tuesday 05 June 2018 (41695) Print this report

This morning we installed a second chiller (same model) and we switched it on to verify it is working properly, as you can see from the comparison with the other one (attached plot).

We will leave this chiller working for one week at least.

Images attached to this comment