Reports 1-1 of 1 Clear search Modify search
AdV-TCS (Point Absorbers Mitigation)
melo - 11:11 Friday 25 September 2026 (69808) Print this report
NI PA thermocamera analysis

After performing an analysis of the WI thermocamera images (see #69785), a similar analysis for the NI was done.

  • Reference Image: The analysis utilizes the thermal image containing the mirror outline and the identified point absorber location (marked at pixel coordinates [X, Y] = [310, 199]), see Fig. 1.
  • Geometric Reconstruction: The mirror boundary (D = 35 cm) was fitted to extract the center in pixel coordinates, yielding C = [346, 204].
  • Perspective Correction: Taking into account the viewport entrance angle of 18.2 degrees from the left (along the decreasing columns axis), the horizontal axis pixel scale is corrected for perspective compression.

Geometric Mapping:

  • X-Axis (Columns): The mirror center is located at column 346, whereas the point absorber is at column 310. This places the point absorber to the left of the center (an offset of -36 pixels in the image, corresponding to approximately -5.7 cm in physical space).
  • Y-Axis (Rows): The mirror center is located at row 204, whereas the point absorber is at row 199. This places the point absorber slightly above the center (an offset of -5 pixels in the image, corresponding to approximately -0.8 cm in physical space).
  • Radial Distance: The total estimated physical distance of the point absorber from the mirror center is approximately 5.8 cm.

Figs. 2 and 3 display the MATLAB output, showing a differential temperature of approximately 0.52 degrees for the point absorber (calculating by subtracting the image of the mirror in CARM_NULL and the image with the ITF in DOWN status). No other strong PA has been spotted. 

Images attached to this report
Search Help
×

Warning

Error

The present report has been modified outside this window. Please check for its integrity in the main page.

Refreshing this page will move this report into drafts.

×