AdV-INJ (General opto-mechanical layout, common parts construction, Installation and pre-commissioning)
derossi, lagabbe, spinicelli - 12:38 Tuesday 19 May 2026 (69110) Print this report
Comment to chiller malfunctioning (69097)

This morning we went in the Eeroom to check the INJ chillers.

- the MAIN CHILLER Omi (which is used for the SL diodes in EEroom, SL crystal in LL and Beam dumps IPC (EIB) + IMC REFL (EIB but not connected) + PMC REFL (LB) + B2 (SIB2) + IPC2 (SIB1 flux blocked)) when switched off was giving the error A05. We changed the filter and switched it on and it was working fine. We switched it off again.

- the NEOVAN HEAD CHILLER Termotek (used for the Neovan head). We switched it on with the circuit closed on itself and this time the pump started (we observed this behaviour also in the past, maybe due to the air which enters while changing the filter)