The test with the QD2 sgnal lasted up to the unlock. I then disabled it by hand as the automation did not know about it, then I added in DRMI_LOCK the reset of this new sensing matrix weight.
At the end of the calibration activity there is clean data from 18:11 UTC for 300 s; then, after a 25-minutes glitch, I changed the BS_TY loop sensor again from QD1 to QD2 at 18:20:03 UTC; clean data afterwards for 20 minutes.
At 18:41:12 UTC (ramptime = 5 s, in ADJUSTING mode) I opened the SR_TY loop, clean (SCIENCE) data for another 20 minutes. I enabled the loop back at 19:04:11 UTC (ramptime= 5 s).