nardecchia, berni - 15:14 Saturday 18 October 2025 (67976) Print this report
WI CO2 main laser malfunctioning

Today around 14:00 LT Francesco called me to report a malfunction of the WI CO2 laser (the one used for DAS).

From the first attached plot it appears that the laser started exhibiting anomalous behavior around 05:00 UTC, with repeated drops in output power followed by recoveries. These oscillations became progressively more frequent in the late morning, eventually leading to several short shutdowns (1–2 minutes each). The last one, the longest, lasted from 12:04 UTC to 12:11 UTC.

After this interruption, the laser restarted by itself without any action from either me or Francesco. The issue seems to have an electrical origin.

Despite the CO2 laser being off for about 7 minutes, the interferometer remained locked, although with an evident overall worsening of the signals. It eventually unlocked at 12:42 UTC.

I followed the lock acquisition again up to CARM NULL 1F and everything went well.

We will keep monitoring the laser behaviour.

Images attached to this report
Comments to this report:
nardecchia, spinicelli, zaza - 20:19 Sunday 19 October 2025 (67983) Print this report

After the manual power cycle reported by Francesco in his previous entry (67979), the WI CO2 laser appeared stable for some hours, until it switched off again and then restarted autonomously.

This happened at 17:38:30 UTC, after which the laser turned back on at 17:39:10 UTC.
A second event occurred at 17:51:50 UTC, with the laser restarting again at 17:52:50 UTC.

Cecilia briefly went to the TCS room to verify the hardware status, but did not observe anything abnormal.

Moreover, Francesco noticed that during every instability of the WI CO2 laser, the PR CHRoCC shows a simultaneous disturbance (see attached plot).
This correlation is quite puzzling, since the CHRoCC power supply is located on a completely different rack from the CO2 electronics.
For this night, there is not much we can do =(

Images attached to this comment
zaza - 20:32 Sunday 19 October 2025 (67984) Print this report

The only anomaly noticed after a second investigation is a soft continuous crackling sound coming only from PR CHROCC K.E.R.T. Other identical devices do not produce this sound.

nardecchia - 10:10 Monday 20 October 2025 (67988) Print this report

I am not sure whether this is relevant, but in correspondence with the power on/off events of the WI CO₂ laser and the glitches in the PR CHRoCC, simultaneous disturbances are visible on the UPS INF_CB_UPS1_RBT (see attached plot).

Images attached to this comment
Tringali, Fiori - 11:29 Tuesday 21 October 2025 (68002) Print this report

By looking at the time interval (Oct 19, 18:20–19:00 UTC) shown in Ilaria’s plot, we observed an increase in noise measured by the accelerometer and the episensor on the bench, coincident with the laser power on/off events (Figure 1). This correspondence is clearly visible around 18:22–18:28 UTC (Figure 2), and less pronounced between 18:40–18:43 UTC (Figure 3). During this period, the ITF was locked.
We also found coincidences in the UPS current and voltage monitors of phase T (Figure 4 - 5). Looking at other ENV sensors in the TCS room and in the CEB hall, no coincidences were observed.

Images attached to this comment
nocera - 9:54 Wednesday 22 October 2025 (68011) Print this report

In an attempt to help disentangle the unexplained coupling between WI CO2 laser hitches and the PR CHRoCC ones, yesterday afternoon we moved the power cable for the latter on a different power line (they were previously sharing the same one)