The SQB1_FI Acl server, used to manage the SQB1 Faraday insulators FI1 and FI2, has been set up on the SQB1_rtpc .This server is running ar 10KHz on the RT_CPU0 .
Expected cabling:
- ADC2378-SN118 (SQB1_ADC_01) - DBOX_SN88 (SQB1_DBOX_LC)
- CH00: Faraday insulator id 1 Temperature probe called SQB1_IR_FI1_TEMP
- CH01: Faraday insulator id 1 current monitoring probe called SQB1_IR_FI1_CURR_MONI
- CH02: Faraday insulator id 2 Temperature probe called SQB1_IR_FI2_TEMP
- CH03: Faraday insulator id 2 current monitoring probe called SQB1_IR_FI1_CURR_MONI
- DAC1955-SN151 (SQB1_DAC_01) - DBOX_SN88 (SQB1_DBOX_LC)
- CH00: Faraday insulator id 1 Peltier cell driving
- CH01: Faraday insulator id 1 Peltier cell driving
The use of SQB1_rtpc’s RT_CPU processors is now
- RT_CPU0 (red rectanble):
- EQB1_HD_CC : loop frequency 10KHz - tas order: 5
- SQB1_FI: loop frequency 10KHz - tas order: 7
- RT_CPU1 (green rectangle):
- SQB1_LC : loop frequency 10KHz - task order: 0
- SQB1_SBE : loop frequency 10KHz - task order: 1
- EDB_QD_MM : loop frequency 10KHz - task order: 2
- RT_CPU2 (blue rectangle):
- EDB_QD_AA : loop frequency 10KHz - tas order: 3
- EQB1_QD : loop frequency 10KHz - tas order: 4
- EQB1_HD_AA: loop frequency 10KHz - tas order: 6