Upon arrival I found ITF relocking. The B1p spot was miscentered respect to the camera, so I moved with the fixed setpoint up to CARM_MC_IR, then I engaged again the floating setpoint.
ITF reached LN3_SQZ but at 22:00 UTC we discovered a problem with TCS_WI_CO2_PWROUT. I contacted TCS expert who fixed the problem (ADJUSTING MODE set).
After a couple of locking attempts ITF locked again in a stable way.
SCIENCE_MODE set at 23:37 UTC.
ITF unlocked 3 times during ACQUIRE_LN1 (TBC) so I decided to wait a bit in DC_READOUT. LN3_SQZ acheived at 04:30 UTC.
SCIENCE_MODE set at 04:31 UTC.
Guard Tours (UTC time)
21:06 - 21:35
23:01 - 23:29
01:02 - 01:31
03:02 - 03:31
DAQ
NewtonNoise
02"15 UTC - NNC_CEB_LEV2 process crashed. I tried to restart it via VPM without success:
Log:
FATAL..-Device V1:NN_CEB_INF_Z_209> No response after IP configuration
WARNING-Device V1:NN_CEB_ACC_Z_226> Writing invalid data with comment INFO_ZEROS, INFO_DATA_INVALID, INNO_DATA_TOO_SHORT, INNO_DATA_LENGTH, SERVICE_TIME_INTERVAL_HIGH, SERVICE_TIME_INTERVAL_SKIPPED, SERVICE_DATA_NO_REQ_LAST_SEC, SERVICE_DATA_OVERWRITTEN,
WARNING-Device V1:NN_CEB_ACC_Z_217> Writing valid data with comment SERVICE_TIME_INTERVAL_HIGH, SERVICE_TIME_INTERVAL_SKIPPED, SERVICE_DATA_NO_REQ_LAST_SEC, SERVICE_DATA_OVERWRITTEN,
TCS
(11-04-2024 21:55 - 11-04-2024 22:20) From remote
Status: Ended
Description: Power_DAS_OUT_is_out_of_range
TCS_WI_CO2_PWROUT at 3.6