bertolini, capocasa, guo - 22:27 Wednesday 12 May 2021 (51769) Print this report
FCIM DDS module problem

Last week, we changed the power supply for the FCIM crate to an independent power supply (see entry 51645)

It didn’t help to solve the jump issue and we found that the coil current monitor channels  became very noisy for marionette vertical coils and mirror coils (pic1). This situation never appeared in the past. So we changed the power supply back to it original one, and the excess of noise disappeared.  

Then we tried to recover the system, starting from the control of the bench but we found that the DDS board was not working anymore. To confirm the problem is the DDS board and not the crate, we plugged the FCIM DDS module to the SQB2 crate, there was still no output signals. So at this moment we are providing the 10kHz signal for LVDT with an external signal generator. (Pic2) Under this condition, we checked again the coil current monitor signals, the jumps still exist (pic3).  We are still suspecting that the electric problems are coming from the crate. A spare crate has been sent from Nikhef this Monday, we expect it to arrive before Friday.

Images attached to this report
Comments to this report:
bertolini - 0:20 Thursday 13 May 2021 (51772) Print this report

After the replacement of the DDS with the function generator, i have re-calibrated the LVDTs by injecting white noise between 0.05-0.2Hz in Z and measuring the transfer function between LVDT-Z and Trillium-Z. Then i have centered the IP using the step motors and close the position loops for the night. In FCIM the new guardian is already implemented.