AdV-DET (Commissioning)
eisenmann, gouaty - 22:05 Saturday 30 March 2019 (45474) Print this report
Crazy values in B7 DC

This evening after the unlock that occured at 17h37 utc, weird values were observed in the SNEB_B7_PD1_DC and SNEB_B7_PD2_DC channels with some negative peaks reaching -18 mW in PD1 and -70 mW on PD2 (see attached figure). The problem lasted up to 18h51 utc on PD1 and 18h55 utc on PD2. Note that Marc sent a command to close the shutter of PD1 at 18h55m02 utc and then reloaded the SNEB_dbox_bench at 19h09m13 utc. Since then the photodiodes signals behaves normally. To be further investigated.

 

Images attached to this report
Comments to this report:
eisenmann, gouaty - 22:32 Saturday 30 March 2019 (45475) Print this report

B7 PD1 and PD2 shutters were closed to check the photodiodes offsets but nothing particular to report.

I then took some time to slightly retune the offsets  which were changed by :

PD1_DC -1.004e-4 mW
PD1_audio -0.687e-4 mW
PD2_DC 3.794e-4 mW
PD2_audio 1.89e-4 mW

 