Reports 1-1 of 1 Clear search Modify search
AdV-PSL (Pre-mode-cleaner block)
cleva - 11:11 Tuesday 11 December 2018 (43985) Print this report
PMC loop status after slave injection

# Status of the PMC lock after the recovery of the proper injection lock condition (43966)

PDH_gain margin = 23 dB
PDH_offset = -100 mV
PDH_pp = 2.45 V pp
RF_Att  = 20 dB
PMC_REFL_DC = 3.4 V (unlocked)
data @ 16h02 LT
we deduce S_PZT ~ 36 MHz/V
see plot for the PDH scan


Rem.: the PDH used to be 2.0 V pp with 9 dB RF_ATT and REFL_DC ~ 2 V (see 40825)
Rem.: that is an increase of the PDH of (2.45/3.4*11dB)/ (2/2) = x 2.48

**********************


# Freq-to-RIN coupling


it becomes possible to anneal the Freq-to-RIN coupling using an offset of -0.2 V on the rampauto perturb input.
It seems this has been attempted in the past weeks with no success (to beat the SSFS_boost correction which couples to PMC_TRA)
 

Images attached to this report
Search Help
×

Warning

Error

The present report has been modified outside this window. Please check for its integrity in the main page.

Refreshing this page will move this report into drafts.

×