Reports 1-1 of 1 Clear search Modify search
AdV-PSL (Laser Bench setup)
Cleva, Kefelian - 10:54 Wednesday 11 April 2018 (41060) Print this report
neoVAN optimization

The purpose of the shift was to check the neoVAN seeder beam profile and try to understand why we encounter some power loss (log entry 40855). This shift happens between 3/04/18 11pm LT and 4/04/18 8am LT and should be associated to the logentry https://logbook.virgo-gw.eu/virgo/?r=40976


# upon arrival 3/04/18- 08h43 LT, we found the PMC_TRA_DC = 1,02 V = 62,9 W (AOM is still connected)

# we got also P_LB_M9 = 94,6 W (P at neoVAN output)

# we made a status of the PSL power ( 3/04/18, both calo zeroed. @ 10:21 LT: Step9_1)
we disconnect the AOM: PMC_TRA_DC moves from 1.022 to 1.065 V = 2,65 W (S_PMC_TRA = 61.68 W/V)
# issue_1: we were rather expecting a raise of 3,5 - 4,65 W (see logentry 40434, second plot).

d        P_LB_M6 Delta_TE02     R_PMC_V_OR    R_PMC_W_OR     R_PMC_Reso     T_PMC     (1-C_R_1)    T_PMC_%
m        mW        mV            V            W            W            W        
0.55    17.2    280            3.47        88.6        22.6        64.3    0.1744        0.9808    Step9_1

Legend:
R_PMC_V/W_OR: OR for OffReso, V/W for Volt or Watt
1-C_R_1 = R_PMC_Reso/R_PMC_W_OR  - Delta_%_TEM02
Delta_%_TEM02 = Delta_TEM02/R_PMC_V_OR
T_PMC_% = (T + R_PMC_Reso)/R_PMC_W_OR

--> the incident power is now 88,6 W instead of 93,7 W in Dec 2017 (see log entry 40855)
--> the PMC transmission is still around 0.98


# we change the current of the neoVAN amplifier to proceed with some calibration
@1206773658: AMP_D2 from 5.3 to 5.2 A (~ 6:54 UTC)
@1206773841: AMP_D2 from 5.2 to 5.3 A
- P_neoVAN decreases by 1,489 %
- P_PMC_TRA increases by 1 %
- P_PMC_REFL decreases by 15 %
--> it moves also the matching wrt PMC: useless for any calibration !?


# we check the seeder beam profile (BP) at the neoVAN input, in transmission of LB_M6 mirror (T~1.084e-3).
we found the BP changed wrt what set on Dec 40260, (logentry 40260, profile "k").
The BP featured a 132 um waist at 0.49 m from LB_L5 and is now
115 um @ 0.50 m from LB_L5 lens.
a change of the telescope focal length from 0.45 m to 0.40 m would explain such a change,
but would also bring the waist closer to the LB_L5 lens by 7 cm, which is not observed.
see plot "k" (old meas.), "l" (4/04/18 new meas.)
-> no clue yet on why the BP_seeder has changed

# we setup a new BP of the seeder (see plot "n")
The BP is now 136/139 um, although not optimal we keep it for the time being (optimal should be 130 um)
the distance(LB_L4 to LB_L5) is now 33 mm (lens to lens) instead of 36 mm.   
LB_L4 is set at 98,7 cm from LB_WP5 (used to be 110,7 cm before)
LB_L5 is set at 0.635 m from neoVAN input (=0.32+.21+.105 m)
it means the seeder waist is at 5,5 cm outside of the neoVAN (nominal is 5cm)

we reduce the neoVAN pump current from 4.7/5.3 A to 4.4/5.3 A
This improves the neoVAN modal content and reduces the PMC mismatch.
-> As a result, although the neoVAN output power is less by 4 W the PMC_TRA is improved by 3-4 W
rem.: the seeder together with the pump is responsible for the neoVAN power change

# Issue
the signal PSL_AMP_DC does not see the neoVAN power change: we suspect the beam is too large on the monitoring PD and it may happen that the HOM are clipped
The modal analysis with the PMC shows that 10% of the neoVAN power is spread on HOM of order 4.
Rem.: outer radius of HOM_order4 ~ 3 x waist


# we made a status of the PSL power before leaving ( 4/04/18, @ 06:50 LT: Step9_3)
d        P_LB_M6 Delta_TE02     R_PMC_V_OR    R_PMC_W_OR     R_PMC_Reso     T_PMC     (1-C_R_1)    T_PMC_%
m        mW        mV            V            W            W            W        
0.55    17.2    280            3.47        88.6        22.6        64.3    0.1744        0.9808    Step9_1
0.635    xx        0.110        3.48        84.6        15.3        68.2    0.148        0.987    Step9_3


## Conclusion:
- the power seems to have decreased from 93.7 W (December) to 88.6 W (April 2018) at the PMC reflection; not clear whether it is due to the seeder size change
- the neoVAN monitor PD is suspicious
- the neoVAN power measured at neoVAN output with a calorimeter does not seems to have changed
- the PMC throughput is still around 98 %
- the work needs to go on with the seeder optimization together with the neoVAN pump optimization

Comments to this report:
cleva - 17:32 Wednesday 11 April 2018 (41072) Print this report

with the plots of the seeder "k", "l", "n"

I 've added the beamp rpofile of the neoVAN in the far field

plot 4 made on 7/03/18 with the "faulty" seeder ( profile "l"), at length focus (F=0.5 m)

plot 5 made on 4/04/18 with the new adjustment on the seeder side (profile "n"), at length focus (F=0.4 m)

Images attached to this comment
Search Help
×

Warning

×