Reports 1-1 of 1 Clear search Modify search
AdV-DAQ (Data Acquisition and Global Control)
rolland - 17:53 Monday 10 July 2017 (38521) Print this report
B1 photodiode signals stored at 100 kHz for drifting line study. Hrec range

The signals SDB2_B1_PD1,PD2_Audio_100kHz have been temporary added in the DAQ for studying the line drifting as requested by Bas.

 

Hrec was giving a very nice horizon of 70 Mpc... but this was due WE actuation relay being set in LN1 mode manually : 'Low Noise 1 [byte=0]' sent to CoilsSbWE' but probably without the associated gain in the  DSP changed accordingly.  Hence the WE actuation response was not the standard one and the reconstructed h(t)  was false.

Search Help
×

Warning

Error

The present report has been modified outside this window. Please check for its integrity in the main page.

Refreshing this page will move this report into drafts.

×