AdV-COM (AdV commissioning (1st part) )mwas - 11:11 Friday 16 June 2017 (38034)
Print this reportComment to SDB1 excitations - many mechanical resonances between 100-1000Hz (38028)I have looked at the data with the SDB1 slow Z excitation. Using the same method as for SDB2 and the SDB1-SDB2 as the position signal works very well.
Figure 1, the observed B1 spectrum and the projected scattered noise. The coupling factor is 3 times larger than for SDB2. So SDB1 is a more critical bench in terms of scattering.
Figure 2, the same for the quite time the noise is about a factor 10 below the sensitivity curve in the 20-80Hz range. So not that far away.
It would be good to look through the viewports with a IR viewer and at the IP cameras to see if there is some obvious source of scattering on SDB1.