Detector Characterisation (Broadband noise)
fiori - 15:30 Friday 17 February 2017 (36561) Print this report
cohorence analysis of tonight 50mins lock

Here the bruco coherence tables over all  ENV channels for tonight lock with 98% dark fringe and DARM on B1p:

Some insights:

  • the big structures (31 Hz bump, intense narrow lines at 54.1Hz, 67.1Hz, 87.1Hz) have NO coherence at all with ENV channels
  • the most evident ENV noise is the 1Hz comb:
    • most coherent channels are (in decreasing order) CEB magnetometer, UPS voltage sniffers, and IPS sniffers: Figure 1: MAG,  UPS,  and IPS
    • from 10Hz to 30Hz some coherence is present also with NEB and WEB UPS voltage (we might suspect something from the cryotrap heater...)  Figures 2 and 3
  • 6Hz peak is coherent with CEB SEISM and Acoustic (Figure 4)
  • 7-10Hz region has some seismic coherence with CEB hall,  a bit stronger with EIB bench (fig. 5)
  • 16.9 Hz seismic line, mostly coherent with TCS WI bench (Figure 6)
  • 24.6 and 24.8 Hz sesmic lines look mostly coherent with EIB and IB (is there some vacuum pump still on there?) Figure 7
  • 28.9 and 31.5 Hz (small) seismic lines are coherenc with CEB SEISM and Acoustic 
  • there are pronounced 50Hz sidebands (48.3, 48.7, 51.3, 51.6 Hz) coherent with CEB magnetometer
  • 200-300Hz some coherence with CEB seism. (Figure 9)
Comments to this report:
bersanetti - 15:40 Friday 17 February 2017 (36566) Print this report

About the structures which show no coherence with other signals: I do not know any possible source for the 31 Hz bump, while the three narrow lines are the UGF servos' lines we inject in the loops in order to track the UGF of the loops.