AdV-INJ (Reference cavity (RFC))
capocasa, carbognani, genin - 17:06 Wednesday 11 January 2017 (36012) Print this report
TCS chillers effect on RFC frequency noise

In order to investigate the contribution of the CO2 laser chiller to the frequency noise we switched them off at 11.48 am local time. In the attached plots there is a comparison of the RFC frequency noise spectrum (with IB angular control noise projections) with the chillers off and on. 

We confirmed that when the chillers are off the 1 Hz comb noise disappears. There is some smaller residual comb noise about 50 Hz to be better investigate.

Chillers will be swiched on again as soon as TCS people need it.