AdV-COM (AdV commissioning (1st part) )
capocasa, barsuglia, degallaix - 12:27 Tuesday 24 May 2016 (33859) Print this report
Estimation of the etalon effect in the arm cavities

We estimated the expected changes in the finesse of the arm cavities due to the etalon effect of the input mirrors.

According to the formula  at pag 54 of Bondu's HDR, the FInesse F (without the etalon effect) is related to the finesse F' (with the etalon effect) by

F'=F*(1-r_AR*cos(b)*(2*chi^2 chi-1)/chi)   where    chi = (-r1 r2*(1-P1))/(1-r1*r2)     

we used the following values

NORTH ARM

r_AR = sqrt (R_AR) with R_AR = 32 ppm         reported in VIR-0544A-14

r1 = sqrt (1-T1)  with T1 = 1.377e-2                 reported in VIR-0544A-14

r2 = sqrt (1-T2) with T2  = 4 ppm                     reported in VIR-269A-15

P1 (arm losses) = 75 ppm

F = pi*sqrt(r1*r2)/(1-r1*r2) = 453 ± 2

We found F'_min = F'(b = 0) =  448  and F'_max (b =pi) = 458

WEST ARM

r_AR = sqrt (R_AR) with R_AR =  58 ppm          reported in VIR-0543A-14

r1 = sqrt (1-T1)  with T1 = 1.375e-2                    reported in VIR-0543A-14

r2 = sqrt (1-T2) with T2  = 4.3 ppm                     reported in VIR-0270A-15

P1 (arm losses) = 75 ppm

F = pi*sqrt(r1*r2)/(1-r1*r2) = 453.7 ± 2.3

We found F'_min = F'(b = 0) =  446.9  and F'_max (b =pi) = 460.5  

In the attached plot these values are compared with the finesse measured from the ringing effect for the noth cavity (see entry 33795).