Detection system (Detection system)
Vajente, Marque - 15:51 Thursday 18 August 2011 (30124) Print this report
B1 setup moved back to External Detection Bench
This morning, we removed the B1 setup from the little bench attached to the detection tower and re-installed it on the External Detection Bench (as it was up to August 9th, see 30068).
After 10 days taking data in this configuration, it was clear that there was no improvement at any frequency due to the shortening of the optical path in air (120cm -> 55cm).
There was however a worsening of the sensitivity around 90Hz. These bumps disappeared when the setup has been moved back to the EDB (plot1).
On the same plot, one can see that another bump around 65Hz, which has always been present, disappeared after this morning intervention (?).
Looking at the coherence with the seismometer attached to the detection tower, there is no clear evidence but some indication that these bumps may be due to backscattered light from the B1 setup.